Z-0269-04

Spacelabs flat panel monitor, Model 91415-A

GE Healthcare / initiated 2003-10-29 / Terminated

Datex-Ohmeda, Inc. began a recall of Spacelabs flat panel monitor, Model 91415-A on . The reason FDA records is "Potential for power supply to fail resulting in failure of the audio (alarm tones) to fail". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0269-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Spacelabs flat panel monitor, Model 91415-A
Product code
DXJ Display, Cathode-Ray Tube, Medical
Reason for recall
Potential for power supply to fail resulting in failure of the audio (alarm tones) to fail.
FDA root cause
Other
Action
A letter dated October 31, 2003 was sent on October 29-31, 2003 to all US customers via FED-EX. An ORDER FORM accompanied the letter for consignees to request a replacement power supply or to indicate they did not have the affected power supply.
Quantity in commerce
2734 units for domestic consignees and 1099 for international consignees
Distribution
The firm distributes to hospitals nationwide. The firm also distributes internationally.
Product codes and lots
Date code range: 0001xx through 0152xx. All defective power supplies fall within the date code of week 1, 2000 and week 52, 2001. The date code can be found on the underside of the power supply, embossed at the top of the text. It will be in the format of 0108A3, which means ''01'' is the year 2001, ''08'' is week 8 and ''A3'' is the line A3.