Z-0182-04
Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number...
Allen Medical Systems, Inc. / initiated 2003-11-06 / Terminated
The OR Group began a recall of Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A on . The reason FDA records is "Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0182-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Allen Medical Systems, Inc.
- Product
- Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A
- Product code
- FPS Board, Bed
- Reason for recall
- Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb
- FDA root cause
- Other
- Action
- The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.
- Quantity in commerce
- 17 units
- Distribution
- Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA
- Product codes and lots
- S/N''S: R00043, R001080, R001092, R001198, R0001341, R001370, R001374, R001366, R001337, R001497, R001478, R001493, R001501, R001529, R001812, R001816, R002023.
FIRST ARTICLE / WEEKLY