Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0182-04

Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number...

Allen Medical Systems, Inc. / initiated 2003-11-06 / Terminated

The OR Group began a recall of Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A on . The reason FDA records is "Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0182-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Allen Medical Systems, Inc.
Product
Ronci Surgical Universal Imaging Extension used with the Ronci Pain Management Table System Catalog Number: R-934-25A
Product code
FPS Board, Bed
Reason for recall
Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb
FDA root cause
Other
Action
The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.
Quantity in commerce
17 units
Distribution
Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA
Product codes and lots
S/N''S: R00043, R001080, R001092, R001198, R0001341, R001370, R001374, R001366, R001337, R001497, R001478, R001493, R001501, R001529, R001812, R001816, R002023.