Z-0142-04

Waste Collection Kit. Label reads in part: ''CUSTOM KIT...MERIT MEDICAL''. Latex free, Sterile if package is...

Merit Medical Systems, Inc. / initiated 2003-10-01 / Terminated / root cause: packaging

Merit Medical Systems, Inc began a recall of Waste Collection Kit. Label reads in part: ''CUSTOM KIT...MERIT MEDICAL''. Latex free, Sterile if package is unopened or undamaged. Part # K10-01047 on . The reason FDA records is "There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0142-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Waste Collection Kit. Label reads in part: ''CUSTOM KIT...MERIT MEDICAL''. Latex free, Sterile if package is unopened or undamaged. Part # K10-01047.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are generated in the seperated trays.
Root cause tag
packaging
FDA root cause
Other
Action
Merit notified sales representatives, domestic and international accounts and dealers immediately by letter dated October 3, 2003. The manufacturer requests that the use of the affected inventory should immediately cease and that on-hand units should be promptly isolated from general inventory for simultaneous inspection by a site representative and a Merit sales representative.
Quantity in commerce
30 units
Distribution
Distributed nationwide. International distribution includes Canada, Japan, Korea, England and France.
Product codes and lots
Lot # A289183