Z-0175-03
Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo...
Siemens Medical Solutions USA, Inc. / initiated 2003-10-16 / Terminated
Siemens Medical Solutions USA, Inc began a recall of Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata... on . The reason FDA records is "Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0175-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient.
- FDA root cause
- Other
- Action
- The recalling firm field representatives are visiting each location to correct the problem through a software upgrade.
- Quantity in commerce
- 118 units
- Distribution
- The products were shipped nationwide to hospitals and medical imagining centers.
- Product codes and lots
- Model numbers 4772906, 7106995, 7104693, 5751438, 7106714, 7104719, 7106425, 7104594, 4772971, and 7106557.