Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0175-03

Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo...

Siemens Medical Solutions USA, Inc. / initiated 2003-10-16 / Terminated

Siemens Medical Solutions USA, Inc began a recall of Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata... on . The reason FDA records is "Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0175-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Magnetic Resonance Imaging. Brand names: Concerto, Concerto Upgrade, Syngo Harmony, Syngo Harmony 087S, Syngo Harmony Upgrade, Syngo Sonata, Syngo Sonata Upgrade, Syngo Symphony, Syngo Symphony Q, and Syngo Symphony Upgrade.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Equipment SAR monitor does not accept new patient data, incorrect RF levels may be administered to the patient.
FDA root cause
Other
Action
The recalling firm field representatives are visiting each location to correct the problem through a software upgrade.
Quantity in commerce
118 units
Distribution
The products were shipped nationwide to hospitals and medical imagining centers.
Product codes and lots
Model numbers 4772906, 7106995, 7104693, 5751438, 7106714, 7104719, 7106425, 7104594, 4772971, and 7106557.