Z-0183-04
Stryker Universal Imaging Extension (OEM) Catalog Number: R--934-TRIO-A1
Allen Medical Systems, Inc. / initiated 2003-11-06 / Terminated
The OR Group began a recall of Stryker Universal Imaging Extension (OEM) Catalog Number: R--934-TRIO-A1 on . The reason FDA records is "Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0183-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Allen Medical Systems, Inc.
- Product
- Stryker Universal Imaging Extension (OEM) Catalog Number: R--934-TRIO-A1
- Product code
- FPS Board, Bed
- Reason for recall
- Imaging table with extension may fail if the table leg is incorrectly positioned and the load exceeds the maximum weight of 400 lb
- FDA root cause
- Other
- Action
- The OR Group forwarded by 2nd day mail on 11/6/03 the repair kit and instructions to the accounts.
- Quantity in commerce
- 2 units
- Distribution
- Nationwide. CA, FL, GA, IA MI, MT, NH, OH, NJ, NY, SC, UT, TX, VA
- Product codes and lots
- Serial Numbers: 217335, 214331.