Z-0210-04
Exacta-Mix 2400 pharmacy compounding system
Baxter Corporation Englewood / initiated 2003-11-07 / Terminated
Baxa Corporation began a recall of Exacta-Mix 2400 pharmacy compounding system on . The reason FDA records is "If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0210-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Corporation Englewood
- Product
- Exacta-Mix 2400 pharmacy compounding system.
- Product code
- NEP System/Device, Pharmacy Compounding
- Reason for recall
- If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber..
- FDA root cause
- Other
- Action
- All users were contacted by phone and fax on 11/7/03 with instructions to safely use the device.
- Quantity in commerce
- 231 systems
- Distribution
- Nationwide. Foreign distribution to UK. No government or military consignees.
- Product codes and lots
- Catalog Number EM2400, all serial numbers.