Z-0206-04
Hill-Rom brand Resident LT bed ambulatory assist bar accessory; accessory P865A
Hill-Rom Inc. / initiated 2003-11-14 / Terminated
Hill-Rom, Inc. began a recall of Hill-Rom brand Resident LT bed ambulatory assist bar accessory; accessory P865A on . The reason FDA records is "The ambulatory assist bar may not lock into position and may come out of place due to incorrect assembly". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0206-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom brand Resident LT bed ambulatory assist bar accessory; accessory P865A.
- Product code
- KMM
- Reason for recall
- The ambulatory assist bar may not lock into position and may come out of place due to incorrect assembly.
- FDA root cause
- Other
- Action
- Customers were advised of the recall by letter dated 11/12/03 and mailed on 11/14/03. A Hill-Rom representative will visit each customer to assure the latching mechanism is properly attached to the bed.
- Quantity in commerce
- 1400
- Distribution
- United States and Canada
- Product codes and lots
- All units distributed between March 1997 and September 2003.