Z-0276-04
Hypoguard Advance Blood Glucose Monitoring System, includes Hypoguard Advance Blood Glucose Meter and 10...
Arkray Factory USA, Inc. / initiated 2003-11-26 / Terminated
Hypoguard USA, Inc. began a recall of Hypoguard Advance Blood Glucose Monitoring System, includes Hypoguard Advance Blood Glucose Meter and 10 Hypoguard Advance Blood Glucose Strips on . The reason FDA records is "When actual blood glucose levels are in the low end of the blood glucose range, the recalled test strips may provide inaccurately high readings". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0276-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arkray Factory USA, Inc.
- Product
- Hypoguard Advance Blood Glucose Monitoring System, includes Hypoguard Advance Blood Glucose Meter and 10 Hypoguard Advance Blood Glucose Strips
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- When actual blood glucose levels are in the low end of the blood glucose range, the recalled test strips may provide inaccurately high readings.
- FDA root cause
- Other
- Action
- The consignees were sent letters dated 11/26/03 which requested that the product be returned.
- Distribution
- The product was distributed to consignees located nationwide in the United States.
- Product codes and lots
- Lots 01143A, 02263A, 03103A, 10222A, 12212A, 12312A, 03183A, 04293A, and 05273A