IMx CA 125 Mode 1 Calibrator, list 7A89-40; Abbott Laboratories, Abbott Park, IL 60064
Abbott Laboratories HPD/ADD/GPRD began a recall of IMx CA 125 Mode 1 Calibrator, list 7A89-40; Abbott Laboratories, Abbott Park, IL 60064 on . The reason FDA records is "The CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0319-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- IMx CA 125 Mode 1 Calibrator, list 7A89-40; Abbott Laboratories, Abbott Park, IL 60064
- Product code
- LTK--
- Reason for recall
- The CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to an outdated control package insert, 39-1642/R3, instead of the current control package insert 34-1047/R4
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Field Correction recall by letter dated 11/25/03 to all customers receiving the affected reagents. The accounts were informed of the incorrect control package insert referenced in the product information letter shipped with each CA 125 product. The letters contained the correct control package insert number, 34-1047/R4, and stated that all CA 125 assay product inserts will be updated shortly to include the revised storage instructions for the CA 125 Calibrators and Controls.
- Distribution
- Nationwide and internationally through Abbott subsidiaries in Argentina, Aruba, Australia, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Peru, Hong Kong, India, Jamaica, Japan, Korea, New Zealand, Pakistan, Puerto Rico, Singapore, Taiwan, Thailand, United Kingdom and Uruguay.
- Product codes and lots
- all lots