Z-0293-04

IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and...

Abbott Medical / initiated 2003-11-23 / Terminated

Abbott Laboratories HPD/ADD/GPRD began a recall of IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and list 2220-20 (100 tests - reagents alone); Abbott... on . The reason FDA records is "Chemotherapeutic concentrations of methotrexate in patient samples may carryover to subsequent samples, which would result in falsely elevated folate values, when performing IMx Folate testing". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0293-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and list 2220-20 (100 tests - reagents alone); Abbott Laboratories, Abbott Park, IL 60064 USA
Product code
CGN--
Reason for recall
Chemotherapeutic concentrations of methotrexate in patient samples may carryover to subsequent samples, which would result in falsely elevated folate values, when performing IMx Folate testing.
FDA root cause
Other
Action
Abbott notified their international affiliates via e-mail on 11/23/03. Important Product Information letters dated 11/24/03 will be sent to the end users by the affiliates, informing them of the possible methotrexate carryover to subsequent samples and falsely elevated folate values.
Quantity in commerce
3,475 packs
Distribution
There was no U.S. distribution. The product was internationally distributed through Abbott subsidiaries in Colombia, Chile, Uruguay, Argentina, Brazil, Canada, England and Australia.
Product codes and lots
all lots