Z-0324-04

ADx Tricyclic Antidepressants Reagent. Distributed under the following label: ADx Tricyclic Antidepressants...

Abbott Diagnostic International, Ltd. / initiated 2003-11-17 / Terminated

Abbott Health Products, Inc. began a recall of ADx Tricyclic Antidepressants Reagent. Distributed under the following label: ADx Tricyclic Antidepressants Reagents, Abbott Laboratories USA, North Chicago... on . The reason FDA records is "Drug product cross reacts with the Tricyclic Antidepressants (TCA) assay". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0324-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Diagnostic International, Ltd.
Product
ADx Tricyclic Antidepressants Reagent. Distributed under the following label: ADx Tricyclic Antidepressants Reagents, Abbott Laboratories USA, North Chicago, IL 60064
Product code
LFH U.V. Spectrometry, Tricyclic Antidepressant Drugs
Reason for recall
Drug product cross reacts with the Tricyclic Antidepressants (TCA) assay.
FDA root cause
Other
Action
The Important Information Letter dated 11/17/03 was mailed to all ADx, TDx/TDxFL and AxSYM TCA reagent customers. The letters were sent to customers by Federal express Priority mail. The letter recommended contacting all customers who have received the reagents within the last 15 months.
Distribution
Distribution was nationwide.
Product codes and lots
Product List number: 9681-55.