Z-0341-04
Baxter Lineo Opticap Disconnect Cap, product codes R5C4599Q (English) and N5C4599Q (dual English/French...
Baxter Healthcare Corporation / initiated 2003-12-05 / Terminated
Baxter Healthcare Corp. began a recall of Baxter Lineo Opticap Disconnect Cap, product codes R5C4599Q (English) and N5C4599Q (dual English/French label); Baxter Healthcare Corporation, Deerfield, IL... on . The reason FDA records is "Reports of loosening of the connections and disconnection of the device". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0341-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Lineo Opticap Disconnect Cap, product codes R5C4599Q (English) and N5C4599Q (dual English/French label); Baxter Healthcare Corporation, Deerfield, IL 60015 USA; 30 units per case
- Product code
- KGZ Accessories, Catheter
- Reason for recall
- Reports of loosening of the connections and disconnection of the device.
- FDA root cause
- Other
- Action
- Baxter sent recall letters dated 12/5/03 to the accounts in the United Kingdom and recall letters dated 12/8/03 to the accounts in Canada, informing them of the reports of loosening of the connections and disconnection of the device, and requesting that they discontinue use of product codes involved and revert the enrolled patients back to their traditional connection system.
- Quantity in commerce
- 4,920 units
- Distribution
- Canada and the United Kingdom
- Product codes and lots
- Product code R5C4599Q, lots H03J02029 and H03J27067 and product code N5C4599Q, lots H03I18083, H03I23018 and H03J29022
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.