Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0297-04

Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott...

Abbott Medical / initiated 2003-12-02 / Terminated

Abbott Laboratories HPD/ADD/GPRD began a recall of Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064 on . The reason FDA records is "The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution used to clean pipettes of the Abbott Commander Flexible Pipetting...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0297-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Commander Flexible Pipetting Center (FPC) Operations Manual, list 6A97-27; Abbott Laboratories, Abbott Park, IL 60064
Product code
JQW--
Reason for recall
The Operations Manual has the wrong Farenheit temperature range listed (102-122) instead of (104-122) for the preparation of the Terg-A-Zyme solution used to clean pipettes of the Abbott Commander Flexible Pipetting Center.
FDA root cause
Other
Action
Abbott sent 'Important Product Information' letters dated 12/2/03 to the end user accounts, advising them of the error in conversion of the Celsius temperature range to Farenheit in the first step in preparing the Terg-A-Zyme cleaning solution. The letter listed the entire corrected solution preparation procedure and advised the accounts to retain the letter with the Abbott FPC Operations Manual for future reference.
Distribution
Nationwide and internationally through Abbott affiliates in Colombia, Brazil, Puerto Rico, Japan, Singapore, Hong Kong and Germany.
Product codes and lots
all revision numbers, all versions