Z-0298-04

Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the...

Ortho-Clinical Diagnostics, Inc. / initiated 2003-11-26 / Terminated / root cause: manufacturing process

Ortho-Clinical Diagnostics began a recall of Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK on . The reason FDA records is "Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0298-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
Product code
JHX Fluorometric Method, Cpk Or Isoenzymes
Reason for recall
Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Customers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.
Quantity in commerce
774 reagent packs
Distribution
738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.
Product codes and lots
Lot 611, Exp. 25 June 2004