Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the...
Ortho-Clinical Diagnostics began a recall of Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK on . The reason FDA records is "Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0298-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
- Product code
- JHX Fluorometric Method, Cpk Or Isoenzymes
- Reason for recall
- Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Customers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.
- Quantity in commerce
- 774 reagent packs
- Distribution
- 738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.
- Product codes and lots
- Lot 611, Exp. 25 June 2004