Z-0326-04

Roche brand OMNI 7 clinical chemistry analyzer; catalog numbers GD0435 and GD0435R

Roche Diagnostics GmbH / initiated 2003-12-15 / Terminated

Roche Diagnostics Corp. began a recall of Roche brand OMNI 7 clinical chemistry analyzer; catalog numbers GD0435 and GD0435R on . The reason FDA records is "If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0326-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche brand OMNI 7 clinical chemistry analyzer; catalog numbers GD0435 and GD0435R.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.
FDA root cause
Other
Action
Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.
Distribution
United States.
Product codes and lots
All units with software version 5.61 or below.