Z-0328-04
Roche brand OMNI 9 clinical chemistry analyzer; catalog numbers GD0475, GD0475R and GD047591
Roche Diagnostics GmbH / initiated 2003-12-15 / Terminated
Roche Diagnostics Corp. began a recall of Roche brand OMNI 9 clinical chemistry analyzer; catalog numbers GD0475, GD0475R and GD047591 on . The reason FDA records is "If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0328-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche brand OMNI 9 clinical chemistry analyzer; catalog numbers GD0475, GD0475R and GD047591.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.
- FDA root cause
- Other
- Action
- Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.
- Distribution
- United States.
- Product codes and lots
- All units with software version 5.61 or below.