Z-0300-04
MVP-10 Pediatric/ Neonatal Volume Ventilator
Bio-Med Devices, Inc. / initiated 2003-11-07 / Terminated
Bio-Med Devices, Inc. began a recall of MVP-10 Pediatric/ Neonatal Volume Ventilator on . The reason FDA records is "Two way relief valve may fail to activate at the low negative pressure of -3cm H2O". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0300-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Med Devices, Inc.
- Product
- MVP-10 Pediatric/ Neonatal Volume Ventilator
- Product code
- MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
- Reason for recall
- Two way relief valve may fail to activate at the low negative pressure of -3cm H2O.
- FDA root cause
- Other
- Action
- Bio Med Devices notified accounts by telephone on 11/7/03, to return the device for replacement and correction.
- Quantity in commerce
- 3 units
- Distribution
- ID, PA, TX
- Product codes and lots
- S/N: 3023031, 3025031, 0957951