MVP-10 Pediatric/ Neonatal Volume Ventilator

Bio-Med Devices, Inc. / initiated 2003-11-07 / Terminated

Bio-Med Devices, Inc. began a recall of MVP-10 Pediatric/ Neonatal Volume Ventilator on . The reason FDA records is "Two way relief valve may fail to activate at the low negative pressure of -3cm H2O". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0300-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Med Devices, Inc.
Product
MVP-10 Pediatric/ Neonatal Volume Ventilator
Product code
MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
Reason for recall
Two way relief valve may fail to activate at the low negative pressure of -3cm H2O.
FDA root cause
Other
Action
Bio Med Devices notified accounts by telephone on 11/7/03, to return the device for replacement and correction.
Quantity in commerce
3 units
Distribution
ID, PA, TX
Product codes and lots
S/N: 3023031, 3025031, 0957951