Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0302-04

Hill-Rom Newborn bassinet; model P247

Hill-Rom Inc. / initiated 2003-12-19 / Terminated

Hill-Rom, Inc. began a recall of Hill-Rom Newborn bassinet; model P247 on . The reason FDA records is "The caster/wheel may come off of the bassinet, causing the bassinet to tip". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0302-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Hill-Rom Newborn bassinet; model P247.
Product code
FMS Bed, Pediatric Open Hospital
Reason for recall
The caster/wheel may come off of the bassinet, causing the bassinet to tip.
FDA root cause
Other
Action
Recall letters dated December 19, 2003 were sent to each customer. Customers were instructed to verify the security of the casters to the base, and to visually check for loose or missing caster screws until Hill-Rom can visit each facility and perform a fix.
Distribution
United States, Australia, Canada, China, France, Germany, Honduras, Israel, Japan, Mexico, Panama, Taiwan, Turkey, United Arab Emirates and United Kingdom.
Product codes and lots
All units distributed between May 1989 and March 2001.