Hill-Rom, Inc. began a recall of Hill-Rom Newborn bassinet; model P248 on . The reason FDA records is "The caster/wheel may come off of the bassinet, causing the bassinet to tip". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0303-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Hill-Rom Newborn bassinet; model P248.
- Product code
- FMS Bed, Pediatric Open Hospital
- Reason for recall
- The caster/wheel may come off of the bassinet, causing the bassinet to tip.
- FDA root cause
- Other
- Action
- Recall letters dated December 19, 2003 were sent to each customer. Customers were instructed to verify the security of the casters to the base, and to visually check for loose or missing caster screws until Hill-Rom can visit each facility and perform a fix.
- Distribution
- United States, Australia, Canada, China, France, Germany, Honduras, Israel, Japan, Mexico, Panama, Taiwan, Turkey, United Arab Emirates and United Kingdom.
- Product codes and lots
- All units distributed between May 1989 and March 2001.