Z-0508-04

TDx/TDxFLx REA Ethanol, list 9545-60; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064

Abbott Medical / initiated 2003-12-05 / Terminated / root cause: labeling and UDI

Abbott Laboratories HPD/ADD/GPRD began a recall of TDx/TDxFLx REA Ethanol, list 9545-60; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064 on . The reason FDA records is "The reagent package inserts for the products contain incorrect information in the Specific Performance Characteristics section. Accuracy by Correlation table". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0508-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
TDx/TDxFLx REA Ethanol, list 9545-60; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064
Product code
DIC--
Reason for recall
The reagent package inserts for the products contain incorrect information in the Specific Performance Characteristics section. Accuracy by Correlation table
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Field Correction recall by Important Product Information letter dated 12/5/03, sent via Federal Express Priority on the same date to all customers receiving the affected reagents. The accounts were informed of the incorrect values for the Specific Performance Characteristics listed in the package inserts of the reagent kits. The letters contained a table listing the incorrect values and the correct values for each of the products, and the users were instructed to replace their product inserts with the enclosed revised package inserts for all product specific reagent kits in their laboratory.
Quantity in commerce
3,222 kits
Distribution
Nationwide and internationally through Abbott subsidiaries in Mexico, Guatemala, Colombia, Chile, Argentina, Brazil, Puerto Rico, San Salvador, Canada, England, Japan, Singapore, Hong Kong, Taiwan, Thailand, Korea, Australia, New Zealand, Honduras, Costa Rica, Bahamas, Panama, Dominican Republic and China.
Product codes and lots
All lots with insert version 69-5787/R4. The corrected insert is version 34-2778/R5.