Prostatron Praktis Type II for Transurethral Microwave Thermotherapy. The Prostatron is a computer-controlled...
Urologix, Inc. began a recall of Prostatron Praktis Type II for Transurethral Microwave Thermotherapy. The Prostatron is a computer-controlled device designed to deliver microwave energy to... on . The reason FDA records is "During treatments, using the Prostatron units, the devices' Temperature Sensing System may change the rectal temperature reading by up to 2 degrees Celsius and may briefly delay updates to the rectal temperature reading". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0345-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Urologix, LLC
- Product
- Prostatron Praktis Type II for Transurethral Microwave Thermotherapy. The Prostatron is a computer-controlled device designed to deliver microwave energy to the prostrate for the treatment of BPH (Benign Prostatic Hyperplasia). The Prostatron utilizes a transurethral microwave antenna, with simultaneous urethral cooling, to heat the prostate. This heating process is regulated through feedback from a series of temperature sensors.
- Product code
- MEQ System, Hyperthermia, Rf/Microwave (Benign Prostatic Hyperplasia),Thermotherapy
- Reason for recall
- During treatments, using the Prostatron units, the devices' Temperature Sensing System may change the rectal temperature reading by up to 2 degrees Celsius and may briefly delay updates to the rectal temperature reading.
- FDA root cause
- Other
- Action
- A letter about the problem, actions to take to avoid the problem, and the proposed software correction was sent to consignees on December 10, 2003. A software upgrade will be performed on all units in March 2004.
- Quantity in commerce
- 146 units
- Distribution
- The product was shipped to consignees located nationwide in the United States and to foreign consignees in Australia, France and Japan.
- Product codes and lots
- Serial numbers PP80023 through PP80040 and PP037 through PP180