Z-0538-04
Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00
Welch Allyn Protocol, Inc. / initiated 2003-12-31 / Terminated
Welch Allyn Protocol, Inc began a recall of Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00 on . The reason FDA records is "Potential failure to audibly alarm due to speaker hardware component failure". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0538-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Welch Allyn Protocol, Inc.
- Product
- Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00
- Product code
- DSI Detector And Alarm, Arrhythmia
- Reason for recall
- Potential failure to audibly alarm due to speaker hardware component failure.
- FDA root cause
- Other
- Action
- The firm sent a letter dated December 31, 2003 instructing their customers to install an audio cable from the CPU to the Flat Panel Monitor to provide redundant speaker capability. The audio cable and installation instructions were included in the letter. The letter also included a Correction Response Form which customers should fill out upon completion of correction.
- Quantity in commerce
- 233 devices
- Distribution
- The firm distributes the product nationwide and internationally.
- Product codes and lots
- CPU Serial Numbers: LT00001-LT000015, LT000017-LT000019, TA00274, TA00415, TA00415-HA, TA00468, TA00554, TA00582, TA 00587, TA00587-HA, TA00639, TA00738, TA00745, TA00854, TA00856, TA00861, TA00880, TA00889, TA00891, TA01200, TA01217, TA01218, TA01253, TA01278, TA01279, TA01286, TA01304, TA01306, TA01327, TA01359, TA01359-HA, TA01387, TA01387-HA, TA01388-TA01398, TA01391HA, TA01400-TA01411, TA01413-TA01423, TA01426-TA01435, TA01433-HA, TA01438--TA01446, TA01448-TA01455, TA01453-HA, TA01457-TA01461, TA01463, TA01463-HA, TA01464, TA01466-TA01469, TA01471-TA01480, TA01473-HA, TA01478-HA, TA01481-HA, TA01482-TA01486, TA01484-HA, TA01485-HA, TA01488, TA01488-HA, TA01490-TA01493, TA01493-HA, TA01499-TA01515, TA01508-HA, TA01509-HA, TA01510-HA, TA01515-HA, TA01517-TA01526, TA01522-HA, TA01523-HA, TA01525-TA01528, TA01530-TA01546, TA01534-HA, TA01535-HA, TA01536-HA, TA01537-HA, TA01544-HA, TA01548-TA01550, TA01552--TA01561, TA01561-HA, TA01563-TA01570, TA01567-HA, TA01568-HA, TA01574.