Bio Rad Laboratories, Inc. began a recall of Platelia Aspergillus EIA on . The reason FDA records is "Potential drug and device interaction with piperacillin/tazobactam (Zosyn) causing positive test result for galactomannan". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0348-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- Platelia Aspergillus EIA
- Product code
- NOM Antigen, Galactomannan, Aspergillus Spp.
- Reason for recall
- Potential drug and device interaction with piperacillin/tazobactam (Zosyn) causing positive test result for galactomannan.
- FDA root cause
- Other
- Action
- Customers were sent a letter, Technical Bulletin and revised package insert on November 6, 2003.
- Quantity in commerce
- 147 kits
- Distribution
- The firm distributed to hospitals and diagnostic laboratories nationwide.
- Product codes and lots
- Lot number: 3F0001 Exp. 2004-05-15