Z-0517-04
LAERDAL Silicone Resuscitator (LSR), Adult size, containing Intake/Reservoir Valve. Sold as Part Nos...
Laerdal Medical a/S / initiated 2003-12-23 / Terminated
Laerdal Medical Corporation began a recall of LAERDAL Silicone Resuscitator (LSR), Adult size, containing Intake/Reservoir Valve. Sold as Part Nos: 870050-Adult Basic w/o Mask in Carton; 870051-Adult... on . The reason FDA records is "Potential for the Flap Valve component of the Intake/Reservoir Valve to become dislodged from its mount, rendering the resuscitator inoperative". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0517-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Laerdal Medical a/S
- Product
- LAERDAL Silicone Resuscitator (LSR), Adult size, containing Intake/Reservoir Valve. Sold as Part Nos: 870050-Adult Basic w/o Mask in Carton; 870051-Adult Complete in Carton; 870052-Adult Standard with Adult Mask 4-5+ in Carton; 870053-Adult Complete in Compact Case; 870055-Adult Complete in Display Case; 875500-Adult Update Kit. Responsible firm on the label: Distributed in the U.S. by Laerdal Medical Corporation, Wappingers Falls, NY, Made in Norway.
- Product code
- BTM Ventilator, Emergency, Manual (Resuscitator)
- Reason for recall
- Potential for the Flap Valve component of the Intake/Reservoir Valve to become dislodged from its mount, rendering the resuscitator inoperative.
- FDA root cause
- Other
- Action
- Letters issued on 1/14/2004. The recall involves the replacement of the Inner Part of the Intake/Reservoir Valve, which is part of the Laerdal Silicone Resuscitator. Recall is to the user level.
- Quantity in commerce
- 2,153 units
- Distribution
- Product was shipped to approximately 264 customers nationwide, and 3 foreign consignees (Mexico, Brazil and Northern Mariana Islands). Product was also sold to a few exporters who shipped units to unknown countries. Customers include distributors and end users. There were 2,403 parts shipped to Laerdal Medical Canada, Ltd., Toronto, Ontario, Canada. That firm is responsible for conducting an independent recall, reportedly with the knowledge of the Canadian government.
- Product codes and lots
- All product shipped after 1/29/2003 is affected. The following lot numbers are the lot numbers for adult, pediatric and preterm size LSR: 0403, 0703, 0803, 0903, 1003, 1103, 1203, 1303, 1403, 1503, 1603, 1703, 1803, 1903, 2003, 2103, 2203, 2303, 2503, 2703, 2903, 3103, 3203, 3303, 3503, 3603, 3703, 3803, 3903, 4003, 4103, 4203, 4303, 4403, 4503, 4603, 4703, 4803, 4903. NOTE: customers were NOT provided with lot numbers. They were instructed to visually examine the Intake/Reservoir valve in order to determine if they have affected product.