Z-0521-04
Magnetom Trio System. Nuclear Magnetic Resonance Imaging Diagnostic Device
Siemens Medical Solutions USA, Inc. / initiated 2003-12-18 / Terminated
Siemens Medical Solutions USA, Inc began a recall of Magnetom Trio System. Nuclear Magnetic Resonance Imaging Diagnostic Device on . The reason FDA records is "tissue roll can become magnetic". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0521-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Magnetom Trio System. Nuclear Magnetic Resonance Imaging Diagnostic Device
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- tissue roll can become magnetic
- FDA root cause
- Other
- Action
- The recalling firm has issued Updated Instructions to replace the stainless steel tissue roll holder. The recalling firm''s representatives are visiting each account to replace the tissue roll holder.
- Quantity in commerce
- 16 units
- Distribution
- The product was shipped to medical facilities in CA, CT, DC, FL, GA, IL, IN, KS, KY, MA, MD, MN, NC, NY, OH, OR, PA, TX, and UT.
- Product codes and lots
- Model #7387074 - Serial #''s: 20501 through 20508, 20510 through 20524, and 20550 through 20570. Model #7106441 - Serial # 20500