Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The...
Datex-Ohmeda Inc began a recall of Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the... on . The reason FDA records is "Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0545-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
- FDA root cause
- Other
- Action
- Recall letters dated February 03, 2004 stated that a firm representative would contact the consignee to schedule a visit to replace software and hardware to correct the problem.
- Quantity in commerce
- 1617 units in the United States
- Distribution
- The product was distributed nationwide to hospitals in the Unites States.
- Product codes and lots
- The affected units in the United States have serial numbers between 40013635 and 40162379.