Z-0537-04
Pump Tube, REF 8170,121; Suction Pump Drain Tube; An accessory to the suction pump designed for use with the...
Richard Wolf Medical Instruments Corp. / initiated 2004-02-02 / Terminated
Richard Wolf Medical Instruments Corp. began a recall of Pump Tube, REF 8170,121; Suction Pump Drain Tube; An accessory to the suction pump designed for use with the Richard Wolf Ultrasound Lithotriptor used to... on . The reason FDA records is "If the drain tube is assembled incorrectly, the suction pump will not operate as intended". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0537-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Richard Wolf Medical Instruments Corp.
- Product
- Pump Tube, REF 8170,121; Suction Pump Drain Tube; An accessory to the suction pump designed for use with the Richard Wolf Ultrasound Lithotriptor used to disintegrate kidney stones, urinary bladder stones and ureter stones; Richard Wolf GmbH, D-75438 Knittlingen Germany
- Product code
- JOL Catheter And Tip, Suction
- Reason for recall
- If the drain tube is assembled incorrectly, the suction pump will not operate as intended.
- FDA root cause
- Other
- Action
- The accounts were telephoned on 2/2/04 and informed of the potential to re-assemble the suction pump drain tubing incorrectly after sterilization. The accounts were sent re-designed replacement tubing sets and were requested to return the old tubing sets to Richard Wolf.
- Quantity in commerce
- 2 tube sets
- Distribution
- Texas and Nebraska
- Product codes and lots
- part number 8170.121 There is no lot number or production code.