Z-0578-04

4-Way Tee with 2 one-way valves, Ref 395945 Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450 Meadowbrook...

Teleflex Medical / initiated 2004-02-03 / Terminated

Rusch Inc began a recall of 4-Way Tee with 2 one-way valves, Ref 395945 Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450 Meadowbrook Parkway, Duluth, GA 30098***. Carton labeling was similar... on . The reason FDA records is "Valve cap may become loose and enter the air circuitry". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0578-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
4-Way Tee with 2 one-way valves, Ref 395945 Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450 Meadowbrook Parkway, Duluth, GA 30098***. Carton labeling was similar to individual unit labeling, except for, ***Qty 50 ea.
Product code
CBP Valve, Non-Rebreathing
Reason for recall
Valve cap may become loose and enter the air circuitry.
FDA root cause
Other
Action
Consignees were notified on 2/3/2004 via certified mail and telephone.
Quantity in commerce
4060 pieces
Distribution
Nationwide and Canada
Product codes and lots