Z-0739-04
Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the...
Asahi Kasei Medical MT Corp. Oita Works / initiated 2001-10-01 / Terminated
Asahi Medical Co Ltd began a recall of Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and... on . The reason FDA records is "Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0739-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Asahi Kasei Medical MT Corp. Oita Works
- Product
- Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 07004 USA
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Customer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.
- FDA root cause
- Other
- Action
- Asahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.
- Quantity in commerce
- 180,000 units estimated
- Distribution
- New Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana
- Product codes and lots
- Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S. The recall is no lot specific.