DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension...
Del Medical Systems Group began a recall of DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension system designed to facilitate a wide range of... on . The reason FDA records is "There is a potentially defective worm gear in the tension adjusting mechanism of the balancer used in the Digital Ceiling Mounted Tube Support System for overhead x-ray units". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0712-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Del Medical, Inc.
- Product
- DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension system designed to facilitate a wide range of procedures; Manufactured by Del Medical Imaging Corporation, Franklin Park, IL 60131 U.S.A.
- Product code
- IYB Tube Mount, X-Ray, Diagnostic
- Reason for recall
- There is a potentially defective worm gear in the tension adjusting mechanism of the balancer used in the Digital Ceiling Mounted Tube Support System for overhead x-ray units.
- FDA root cause
- Other
- Action
- Del Medical sent a Field Correction Recall letter dated 2/13/04 to all dealers/assemblers receiving DCTM''s during the affected time frame, alerting them to potential defective worm gear. The accounts were asked to locate the systems listed in the attachment to the letter, and alert the end users to take the units out of service until they can be inspected. Follow-up letters were sent via registerd mail on 2/19/04, reiterating the 2/13/04 letter and providing the accounts with instructions for inspection of the balancers. If the worm gear fails the test, the dealers were instructed to contact Del Medical for replacement of the balancer and assure that the unit is not placed back in service until the balancer has been replaced.
- Quantity in commerce
- 57 units
- Distribution
- Nevada, Ohio, Tennessee, Nebraska, Arizona, Maryland, Wisconsin, Florida, Texas, Kansas, Oklahoma, Illinois, Colorado, California, Minnesota, Michigan, Alabama, Rhode Island and internationally to Canada, Panama and Pakistan
- Product codes and lots
- Model DCTM, serial numbers 00182-0303, 00337-0303, 00901-0503, 00948-0503, 02082-0703, 02083-0703, 02152-0703, 02153-0703, 02187-0703, 02188-0703, 02289-0703, 02460-0803, 02492-0803, 02493-0803, 02581-0803, 02730-0803, 02805-0803, 02809-0903, 02810-0903, 02811-0903, 02812-0903, 02866-0903, 02918-0903, 02989-0903, 02990-0903, 02991-0903, 03147-0903, 03148-0903, 03149-0903, 03457-1003, 03458-1003, 03544-1003, 03545-1003, 03624-1003, 03625-1003, 03756-1003, 03757-1003, 03888-1103, 03956-1103, 04085-1103, 04122-1103, 04154-1103, 04236-1103, 04248-1103, 04264-1103, 04371-1203, 04433-1203, 04439-1203, 04480-1203, 04529-1203, 04572-1203, 04634-1203, 04659-1203, 04709-1203, 00032-0104, 00085-0104, WG0031-0102-R