Z-0579-04

RFMD T-Piece With 2 one-way valves & tubing, Ref 395940, Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450...

Teleflex Medical / initiated 2004-02-03 / Terminated

Rusch Inc began a recall of RFMD T-Piece With 2 one-way valves & tubing, Ref 395940, Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450 Meadowbrook Parkway, Duluth, GA 30098***. Carton... on . The reason FDA records is "Valve cap may become loose and enter the air circuitry". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0579-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
RFMD T-Piece With 2 one-way valves & tubing, Ref 395940, Lot***, SINGLE USE ONLY, Qty 1 ea, RUSCH 2450 Meadowbrook Parkway, Duluth, GA 30098***. Carton labeling was similar to individual unit labeling, except for, ***Qty 50 ea.
Product code
CBP Valve, Non-Rebreathing
Reason for recall
Valve cap may become loose and enter the air circuitry.
FDA root cause
Other
Action
Consignees were notified on 2/3/2004 via certified mail and telephone.
Quantity in commerce
13560 pieces
Distribution
Nationwide and Canada
Product codes and lots
Catalog No. 395940, Lot numbers: 108273, 109600, 111314, 111229, 113984, 113932, 113931, 112991, 112514, 111588, 111589, 111558, 111557, 117867, 117866, 116471, 116163, 115616, 114992, 116865, 118714, 118055, 118054, 118053, 119421, 120628, 120521, 120021,120741, 120742, 122619, 122621, 122620, 123017, 123016, 123837, 126217, 125471, 125117, 124069, 126626, 126625, 127052, 127051, 128481, 128322, 127549, 127550, 129360, 129109, 129991, 130186, 130187.