Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0731-04

Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model #...

Boston Scientific Corporation / initiated 2004-02-10 / Terminated

Boston Scientific Corporation began a recall of Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model # M001452150, Catalog #45-215 on . The reason FDA records is "Reports of catheter separation/fracture after implantation resulting in distal migration of the catheter". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0731-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model # M001452150, Catalog #45-215
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
Reports of catheter separation/fracture after implantation resulting in distal migration of the catheter.
FDA root cause
Other
Action
Letters dated 2/9/04 with instructions to return current inventory . Also included is information concerning implanted ports.
Quantity in commerce
786 units
Distribution
Shipments were nationwide to 158 medical facilities nationwide.
Product codes and lots
866561, 876068, 885270, 885444, 886041, 886398, 887738, 888993, 889702, 890163, 903721