Z-0731-04
Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model #...
Boston Scientific Corporation / initiated 2004-02-10 / Terminated
Boston Scientific Corporation began a recall of Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model # M001452150, Catalog #45-215 on . The reason FDA records is "Reports of catheter separation/fracture after implantation resulting in distal migration of the catheter". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0731-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Vaxcel Port with PASV Valve, 8 Fr polyurethane catheter, Mini Titanium port, Implantable Port System, Model # M001452150, Catalog #45-215
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Reports of catheter separation/fracture after implantation resulting in distal migration of the catheter.
- FDA root cause
- Other
- Action
- Letters dated 2/9/04 with instructions to return current inventory . Also included is information concerning implanted ports.
- Quantity in commerce
- 786 units
- Distribution
- Shipments were nationwide to 158 medical facilities nationwide.
- Product codes and lots
- 866561, 876068, 885270, 885444, 886041, 886398, 887738, 888993, 889702, 890163, 903721