Z-0764-04
The BD Blood Glucose Monitoring Systems identified as the BD Latitude, and BD Logic. Becton Dickinson also...
Becton Dickinson & Co / initiated 2004-01-29 / Terminated
Becton Dickinson & Company began a recall of The BD Blood Glucose Monitoring Systems identified as the BD Latitude, and BD Logic. Becton Dickinson also distributes the Paradigm Link Blood glucose Monitor... on . The reason FDA records is "39 lots of BD Test strip may produce an increased frequency of E-3 messages when used during blood glucose testing". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0764-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- The BD Blood Glucose Monitoring Systems identified as the BD Latitude, and BD Logic. Becton Dickinson also distributes the Paradigm Link Blood glucose Monitor which is sold exclusively for Medtronic Mini Med. US Market BD Blood Glucose Monitoring System kit catalog Numbers: 322025 (BD Logic Blood Glucose Monitor Durable Medical Equipment consignees); 322050 (BD Latitude Diabetes Management System); 322051 (BD Logic Blood Glucose Monitor); 32205175 ( BD Logic Blood Glucose Monitor); 322200 (Paradigm Link Blood Glucose Monitor - clear); 322201 ( Paradigm Link Blood Glucose Monitor - smoke); 322202 (Paradigm Link Blood Glucose Monitor - blue); 322203 (Paradigm Link Blood Glucose Monitor - purple). Canadian market BD Blood Glucose Monitoring System kit catalog numbers: 322000 (BD Latitude Diabetes Management System); 322001 (BD Logic Blood Glucose Monitor).
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- 39 lots of BD Test strip may produce an increased frequency of E-3 messages when used during blood glucose testing.
- FDA root cause
- Other
- Action
- Trade and health care professionals were sent recalled letters dated 1/29/2004.
- Quantity in commerce
- 207092
- Distribution
- The product was shipped nationwide to 28 Wholesalers, 15 retail pharmacies and pharmacy chains, 18 managed markets/Mail order facilities, and 616 health care professionals. There are 29 Canadian consignees.
- Product codes and lots
- US Market BD Logic Lot Numbers: 2057320; 2057325; 2507327; 2057361; 2064338; 2064339; 2064341; 2071322; 2071351; 2078324; 2078326; 2078351; 2085324; 2085339; 2085351; 2085360; 2092322; 2092341; 2099322; 2099360; 2106325; 2113325; 3057022; 3057136; 3057147; 3064143; 3064147; 3064148; 3071015; 3071122; 3071125; 3071127; 3071170; 3078023; 3078037; 3078126; 3078135; 3078136; 3078141; 3078142; 3078174; 3085127; 3085139; 3085142; 3085149; 3085161; 3085174; 3092021; 3092122; 3092147; 3092174; 3092176; 3092177; 3099020; 3099147; 3099153; 3099171; 3099174; 3099176; 3099178; 3106147; 3106162; 3113141; 3120176; 3127176. US Market BD Latitude Lot Numbers: 2057360; 2064360; 2071352; 2078352; 2078353; 2085353; 2092354; 2092360; 2099354; 2099357; 2106357; 2113357; 3057020; 3057021; 3057052; 3064021; 3064055; 3064056; 3064064; 3071021; 3071064; 3085090; 3092106; 3106022. US Market Paradigm Link Blood Glucose Monitors Lot Numbers (co-Branded BD Logic): 3057049; 3057051; 3057055; 3057056; 3057058; 3057063; 3057064; 3057070; 3057071; 3057073; 3057098; 3057115; 3057120; 3064045; 3064058; 3064059; 3064077; 3064091; 3064097; 3071049; 3071056; 3071072; 3071076; 3071078; 3071080; 3071090; 3071104; 3071112; 3071120; 3078044; 3078045; 3078051; 3078064; 3078069; 3078077; 3078090; 3078101; 3078106; 3085055; 3085056; 3085079; 3085092; 3085093; 3085105; 3085106; 3085108; 3092044; 3092069; 3092085; 3092101; 3092104; 3099055; 3099083; 3099101; 3099108; 3099125; 3106072; 3106078; 3106097; 3106100; 3106118; 3113042; 3113069; 3113107; 3120104; 3127055; 3127056; 3127059; 3127076; 3134090; 3134107; 3141069; 3141112; 3155100. US Market Paradigm Link (clear) Lot Numbers: 3057199; 3064189; 3064246; 3071178; 3085198; 3085281; 3085282; 3092188; 3120177. US Market Paradigm Link (smoke) Lot Numbers: 3057183; 3057184; 3057197; 3057198; 3057227; 3057241; 3064204; 3064255; 3064267; 3071196; 3071199; 3071203; 3071245; 3071255; 3078202; 3085184; 3085287; 3092287. US Market Paradigm Link (Blue) Lot Numbers: 3057189; 3057195; 3057233; 3064196; 3064197; 3064262; 3071232; 3071289; 3078196; 3078294; 3085181; 3085188; 3085261; 3085293; 3113178; 3113182. US Market Paradigm Link (purple) Lot Numbers: 3057188; 3057196; 3057280; 3064192; 3064195; 3071205; 3071279; 3078184; 3078204; 3078232. Canadian BD Latitude Lot Numbers: 3057041; 3057042; 3057043. Canadian BD Logic Lot Numbers: 3057203; 3057205; 3057224; 3057234; 3064199; 3064203; 3064205; 3064224; 3071202; 3071204; 3071237; 3078198; 3078199; 3078205; 3085202; 3099204; 3120198.