Z-0721-04

VITEK GPS-107Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per...

bioMerieux, Inc. / initiated 2004-01-28 / Terminated

bioMerieux Inc began a recall of VITEK GPS-107Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is... on . The reason FDA records is "Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0721-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
VITEK GPS-107Gram Positive Susceptibility Card, Catalog #V4368, for in-vitro diagnostic use, 20 cards per package. The responsible firm on the label is bioMerieux, Inc., Durham, NC.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
FDA root cause
Other
Action
The firm issued recall letters dated 1/28/04 via regular mail requesting remaining inventories be destroyed and explaining the method which the customer can use to confirm the cards were not misreported. A 'Second Notification' recall letter was issued dated 3/9/04.
Quantity in commerce
1,988/20-card boxes
Distribution
Distribution was made nationwide to medical facilities, which includes military and government consignees. Distribution was also made to foreign consignees located in Colombia, Costa Rica, and China.
Product codes and lots
Lot #M83X, Exp. 4/15/05