Z-0728-04

Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components...

Eastman Kodak Company / initiated 2004-02-20 / Terminated

Eastman Kodak Co began a recall of Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture... on . The reason FDA records is "Defective component in overhead tube support balancer may cause the suspended load to fall". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0728-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Defective component in overhead tube support balancer may cause the suspended load to fall.
FDA root cause
Other
Action
2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced.
Quantity in commerce
3 units
Distribution
Amsterdam, China and Kuwait.
Product codes and lots
Serial Numbers affected: 4193-5702, 4193-5704, 4193-5705, 4193-5706, 4193-5708