Z-0728-04
Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components...
Eastman Kodak Company / initiated 2004-02-20 / Terminated
Eastman Kodak Co began a recall of Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture... on . The reason FDA records is "Defective component in overhead tube support balancer may cause the suspended load to fall". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0728-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- Kodak DirectView DR 7100 System, Catalog #133-6049, a digital radiology system. The major system components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- Defective component in overhead tube support balancer may cause the suspended load to fall.
- FDA root cause
- Other
- Action
- 2/20/2004: Field representatives notified the customer in China that the unit should not be used until the OTS Balancer component can be repaired/replaced.
- Quantity in commerce
- 3 units
- Distribution
- Amsterdam, China and Kuwait.
- Product codes and lots
- Serial Numbers affected: 4193-5702, 4193-5704, 4193-5705, 4193-5706, 4193-5708