Z-0808-04
1757-1: Compressor/Distractor Instrument (aka Plate Distractor Forceps), and product literature. Devices are...
Spinal Concepts, Inc. / initiated 2004-03-01 / Terminated / root cause: design
Spinal Concepts, Inc began a recall of 1757-1: Compressor/Distractor Instrument (aka Plate Distractor Forceps), and product literature. Devices are manufactured by Spinal Concepts, Inc., An Abbott... on . The reason FDA records is "Design has insufficient articulating distance between the two component plates to prevent disassociation in vivo". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0808-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spinal Concepts, Inc.
- Product
- 1757-1: Compressor/Distractor Instrument (aka Plate Distractor Forceps), and product literature. Devices are manufactured by Spinal Concepts, Inc., An Abbott Laboratories Company
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Design has insufficient articulating distance between the two component plates to prevent disassociation in vivo.
- Root cause tag
- design
- FDA root cause
- Other
- Action
- The firm initiated the recall via telephone on 03/01/2004 to all distributors followed up by recall letter on 03/05/2004.
- Quantity in commerce
- 25 (1757-1)
- Distribution
- Nationwide
- Product codes and lots
- All lots