Z-0787-04
Philips Brilliance CT Scanner System, 6, 10 & 16 slice images
Philips North America LLC / initiated 2004-03-08 / Terminated
Philips Medical Systems (Cleveland) Inc began a recall of Philips Brilliance CT Scanner System, 6, 10 & 16 slice images on . The reason FDA records is "The CT couch may move downward to its lower limit without command". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0787-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Brilliance CT Scanner System, 6, 10 & 16 slice images.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The CT couch may move downward to its lower limit without command.
- FDA root cause
- Other
- Action
- The firm contacted their Field Service Personnel by letter on 3/8/2004. The letter instructs the field personnel to immediately install the corrective firmware kits (from Siemens) on all affected units in the field.
- Quantity in commerce
- 97 units
- Distribution
- The systems were installed at medical facilities located nationwide and worldwide.
- Product codes and lots
- Serial Numbers: 2555 to 2568; 3335 to 3392; 9701 to 9725.