Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0783-04

BD L-Cath NN (24ga x 30cm), L-Cath Catheter System

Becton Dickinson Infusion Therapy Systems Inc. / initiated 2004-03-09 / Terminated

Becton Dickinson Infusion Therapy began a recall of BD L-Cath NN (24ga x 30cm), L-Cath Catheter System on . The reason FDA records is "Catheters have stylets extending past the catheter tip, which, if not noticed could cause injury upon insertion". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0783-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Product
BD L-Cath NN (24ga x 30cm), L-Cath Catheter System
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
Catheters have stylets extending past the catheter tip, which, if not noticed could cause injury upon insertion.
FDA root cause
Other
Action
All consignees were notified by telephone on 03/09/2004, followed by a fax letter.
Quantity in commerce
5690 units
Distribution
Nationwide. Foreign distribution to Australia, Canada and Japan.
Product codes and lots
REF 384522, Lot numbers 2226035, 3274339, 3240808, 3294199, 3365138, 3220423, 3192307, 3073801. Added on 6/10/2004:-- 2252516, 2268877.