Z-0826-04

Stat Profile Critical Care Xpress (CCX) Analyzer Catalog Number: 35942

Nova Biomedical Corporation / initiated 2004-03-30 / Terminated / root cause: manufacturing process

Nova Biomedical Corporation began a recall of Stat Profile Critical Care Xpress (CCX) Analyzer Catalog Number: 35942 on . The reason FDA records is "Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0826-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nova Biomedical Corporation
Product
Stat Profile Critical Care Xpress (CCX) Analyzer Catalog Number: 35942
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Root cause tag
manufacturing process
FDA root cause
Other
Action
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
Quantity in commerce
35 units
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Product codes and lots
Domestic S/N: Y01C03200 Y01C03130 Y01104030 Y01104040 Y01104100 Y01104110 Y01304150 Y01304120 Y01104150 Y01104130 Y01104050 Y01104190 Y01104160 Y01104220 Y01204020 Y01204100 Y01204010 Y01204130 Y01204150 Y01204160 Y01703120 Y01304040 Y01304010 Y01304110  International S/N: Inst. S/N Y01204200 Y01304030 Y01304130 Y01304140 Y01304090 Y01304100 Y01104140 Y01204070 Y01204170 Y01104120 Y01C03150