ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and...

Abbott Medical / initiated 2004-03-19 / Terminated / root cause: software

Abbott Laboratories, Inc began a recall of ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product... on . The reason FDA records is "When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0818-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product is manufactured and distributed by Abbott Laboratories Diagnostic Division, Irving, TX 75038
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
Root cause tag
software
FDA root cause
Other
Action
Customers were notified by letter on March 23, 2004 to contact their LIS vendor and to reconfigure the ASTM Field 10.1.4 to allow a maximum character length of 20 characters.
Quantity in commerce
101
Distribution
Nationwide and to the following foreign countries: Chile, Canada, Germany, Taiwan, Australia, Signapore
Product codes and lots
The RS-232 Manual is not controlled by control or lot numbers.