Z-0828-04
Stat Profile Critical Care Xpress CCX (R) Analyzer Catalog Number: 37562 R= Refurbished
Nova Biomedical Corporation / initiated 2004-03-30 / Terminated / root cause: manufacturing process
Nova Biomedical Corporation began a recall of Stat Profile Critical Care Xpress CCX (R) Analyzer Catalog Number: 37562 R= Refurbished on . The reason FDA records is "Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0828-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nova Biomedical Corporation
- Product
- Stat Profile Critical Care Xpress CCX (R) Analyzer Catalog Number: 37562 R= Refurbished
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
- Quantity in commerce
- 3 units
- Distribution
- CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
- Product codes and lots
- Serial Numbers: Inst. S/N Y01902211 Y01203021 Y01B02101