Z-0902-04

Renalsoft software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085

Baxter Healthcare Renal Div / initiated 2004-04-12 / Terminated / root cause: software

Baxter Healthcare Renal Div began a recall of Renalsoft software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085 on . The reason FDA records is "There are software anomalies present that may increase the possibility of a medication error if the dialysis clinic software is used to track, record, or coordinate the administration of medications". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0902-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Renal Div
Product
Renalsoft software; Baxter Healthcare Corporation, McGaw Park, Illinois 60085
Product code
KPF--
Reason for recall
There are software anomalies present that may increase the possibility of a medication error if the dialysis clinic software is used to track, record, or coordinate the administration of medications.
Root cause tag
software
FDA root cause
Other
Action
Baxter sent Urgent Device Correction letters dated 4/12/04 to the attention of the Dialysis Unit Administrator. Separate letters were sent to the Renalsoft customers and to the Renal Software Suite customers. The letters described the software anomalies associated with the software and provided interim workaround instructions until permanent corrective actions to these anomalies can be identified.
Quantity in commerce
16 units
Distribution
Nationwide
Product codes and lots
product code RENLSFT, all software versions