Z-0812-04

Medtronic GEM DR Implantable Cardioverter Defibrillators, Model 7271

Medtronic, Inc. / initiated 2004-04-02 / Terminated

Medtronic Inc. Cardiac Rhythm Managment began a recall of Medtronic GEM DR Implantable Cardioverter Defibrillators, Model 7271 on . The reason FDA records is "Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0812-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic GEM DR Implantable Cardioverter Defibrillators, Model 7271
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Some of the devices with suspect capacitors (high voltage capacitors associated with prior notifications in 1999 and in 2000) have had unexpected charge circuit time-outs or charge circuit inactive conditions as the battery voltage nears the level for Elective Replacement of the devices.
FDA root cause
Other
Action
A letter, dated April 2004, was sent on April 5, 2004 to physicians following patients who had the defibrillators implanted. The letter gave recommendations to physicians for selective replacement of some of the implanted devices and for avoiding the problem. Physicians are being visited by sales representatives to verify notification.
Quantity in commerce
250 units (34 units in the United States)
Distribution
Nationwide in the United States and worldwide.
Product codes and lots
Manufacturing Dates from May 1997 to August 1998; and  Use Before Dates from November 1998 to February 2000.