Image Navigator feature in the MagicView 300 VA42A or VA42B Software. Siemens Part Number: 7502003 and 8463163
Siemens Medical Solutions USA, Inc began a recall of Image Navigator feature in the MagicView 300 VA42A or VA42B Software. Siemens Part Number: 7502003 and 8463163 on . The reason FDA records is "Recalling firm became aware of a potential problem through the submission of a product complaint to the firm reporting the Image Navigator feature in the MagicView 300 VA42A or VA42B software are not being imported...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1095-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Image Navigator feature in the MagicView 300 VA42A or VA42B Software. Siemens Part Number: 7502003 and 8463163
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Recalling firm became aware of a potential problem through the submission of a product complaint to the firm reporting the Image Navigator feature in the MagicView 300 VA42A or VA42B software are not being imported accurately to certain CT or MR images.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The recalling firm has issued a letter to their Siemens Service Engineers to visit each affected sites, in order to correct the software problems. Additionally, a letter advising of the software problems was also sent to the Radiology Director at the affected sites, informing them of the 'Customer Safety Advisory' and the pending visit of the Seimens Service Engineers to perform a software update. Along with these letters, is a set of instructions detailing the potential risks and that a representative from the recalling firm will be contacting the medical facilities.
- Distribution
- Nationwide. The product was shipped to various medical facilities, in the following states, AK, GA, FL, IA, MA, MI, MO, NH, NJ, NY, NV, OK, PA, TX, and WI.
- Product codes and lots
- All software associated with Image Navigator feature in Magic View