Z-0853-04
GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG
GE Healthcare / initiated 2004-03-30 / Terminated
General Electric Med Systems began a recall of GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG on . The reason FDA records is "The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0853-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG
- Product code
- EAM Screen, Intensifying, Radiographic
- Reason for recall
- The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
- FDA root cause
- Other
- Action
- Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
- Quantity in commerce
- 1236 devices
- Distribution
- The units were distributed nationwide in the United States and worldwide.
- Product codes and lots
- All serial numbers