Z-0905-04
Coulter LH750 Hematology Analyzer. Part Number 6605632
Siemens Healthcare Diagnostics Inc. / initiated 2004-03-31 / Terminated
Beckman Coulter Inc began a recall of Coulter LH750 Hematology Analyzer. Part Number 6605632 on . The reason FDA records is "If a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0905-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Coulter LH750 Hematology Analyzer. Part Number 6605632.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- If a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once they are received at the workstation.
- FDA root cause
- Other
- Action
- A Product Corrective Action Letter was mailed to all customers who currently have the instrument in thier lab. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occuring. The letter was sent by US Mail on March 31, 2004.
- Quantity in commerce
- 981
- Distribution
- Nationwide and Canada
- Product codes and lots
- Software versions 2A2 and 2B1.