Z-0918-04

Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL...

MRL, Inc., A Welch Allyn Company / initiated 2004-04-26 / Terminated

MRL, Inc., A Welch Allyn Company began a recall of Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL, Inc., A Welch Allyn Company, 100 Asbury Drive, Buffalo... on . The reason FDA records is "A faulty component has been identified which could result in an intermittent shock button (button #3) failure on the AED10 defibrillator". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0918-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
MRL, Inc., A Welch Allyn Company
Product
Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL, Inc., A Welch Allyn Company, 100 Asbury Drive, Buffalo Grove, Illinois 60089; Model AED10, part numbers 970300 (defibrillator) and 970302 (defibrillator with soft shell carrying case)
Product code
MKJ--
Reason for recall
A faulty component has been identified which could result in an intermittent shock button (button #3) failure on the AED10 defibrillator.
FDA root cause
Other
Action
Urgent Medical Device Recall letters dated 4/26/04 were sent to the end users via certified mail on 4/26/04. The letters informed the users of the intemittent shock (fire or 3) button failure on the AED 10. The letters included a list of the affected serial numbers, and requested the users to call Welch Allyn MRL Technical Support as soon as possible at 1-800-462-0777 for a replacement AED 10 and to answer any questions. A replacement AED 10 will be sent to the end users within 10 days, and the users were requested to return their AED 10 in the same shipping box accompanying the replacement unit. A follow-up letter will accompany the replacement unit with instructions on how to verify the affected units, packing instructions for the return of the affected units to Welch Allyn MRL, and a correction response form to be completed and enclosed with the returned unit.
Quantity in commerce
460 units
Distribution
Nationwide, and internationally to the United Kingdom, Ireland, Singapore, France and South Africa.
Product codes and lots
part numbers 970300 and 970302; serial numbers 4613, 4888/4891, 4897, 6369/6386, 6396/6704, 6706, 6709/6743, 6745/6803, 6806, 6807, 6809, 6811, 6813/6832, 6836/6838, 6853/6855, 6912, 6914, 6917, 6921, 6922, 6924/6926