Port-A-Cath II Low Profile Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter...
Deltec, Inc began a recall of Port-A-Cath II Low Profile Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter, 2.6mm O.D. x 1.6mm I.D., 8.5 Fr Introducer Set, REF... on . The reason FDA records is "Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0867-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Port-A-Cath II Low Profile Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter, 2.6mm O.D. x 1.6mm I.D., 8.5 Fr Introducer Set, REF 21-4071-24
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- A recall letter dated March 26, 2004 gave customers the option of returning the products or using the products if the customer's procedures do not require sterility of the outer surface of the inner tray.
- Distribution
- The recalled products were shipped nationwide in the United States and worldwide.
- Product codes and lots
- Lots M26963, M27058, and M27120