Z-0880-04

Colleague 3 Triple Channel Volumetric Infusion Pumps, product code 2M8153 and 2M8153K (United Kingdom)...

Baxter Healthcare Corporation / initiated 2004-04-13 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Colleague 3 Triple Channel Volumetric Infusion Pumps, product code 2M8153 and 2M8153K (United Kingdom); Baxter Healthcare Corporation, Deerfield, IL 60015... on . The reason FDA records is "Ink deterioration of the letters, numbers and symbols of the programming and pump channel keypads, which may impact the legibility of the keys over time". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0880-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Colleague 3 Triple Channel Volumetric Infusion Pumps, product code 2M8153 and 2M8153K (United Kingdom); Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Singapore; monochrome display screen
Product code
FRN Pump, Infusion
Reason for recall
Ink deterioration of the letters, numbers and symbols of the programming and pump channel keypads, which may impact the legibility of the keys over time.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Baxter sent urgent device correction letters dated 4/13/04 to the direct accounts to the attention of the Director of Biomedical Engineering on the same date. The letters informed the accounts of the ink deterioration of the letters, numbers and symbols on the Colleague Volumetric Infusion Pump programming and pump channel keypads, and informed them that over time such ink deterioration may impact legibility of the keys. The letters provided a listing of model numbers and serial numbers of pumps at their facility which may have potentially affected keypads. The accounts were given a sample of the overlay to be applied over the programming keypad and installation instructions as an interim fix while replacement keypads are obtained and installed as the pumps are processed through their depot repair centers. The accounts were requested to complete the attached reply form and return it via fax to Baxter. The completed form acknowledges receipt of the recall letter, verifies the model numbers and serial numbers of the pumps at the facility and indicates the number of overlays needed. Any questions were directed to Baxter at 1-800-843-7867. On 6/9/05, Baxter sent Urgent Device Correction letters to the accounts, to the attention of the Director of Biomedical Engineering, advising them that the replacement pump programming and pump channel keypads were now available.
Distribution
Nationwide, and internationally to Australia, the Bahamas, Belgium, Brazil, Canada, Colombia, El Salvador, Guatemala, Honduras, Israel, Lebanon, Mexico, New Zealand, Panama, the Philippines, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates and the United Kingdom.
Product codes and lots
product code 2M8153 - serial numbers 12110481CT and above; product code 2M8153K - serial numbers 12120001TK and above