Z-0957-04
MicroSTAAR Injector, MSI-PF Packed ten per box
Staar Surgical AG / initiated 2004-04-21 / Terminated
Staar Surgical Co Inc began a recall of MicroSTAAR Injector, MSI-PF Packed ten per box on . The reason FDA records is "Cracking of the distal end of the injector occurs during locking of the cartridge into the nose cone of the injector". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0957-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Staar Surgical AG
- Product
- MicroSTAAR Injector, MSI-PF Packed ten per box.
- Product code
- KYB Lens, Guide, Intraocular
- Reason for recall
- Cracking of the distal end of the injector occurs during locking of the cartridge into the nose cone of the injector.
- FDA root cause
- Other
- Action
- Recall will be implemented by visits by company representatives.
- Quantity in commerce
- 124 boxes, 10 per box.
- Distribution
- Nationwide
- Product codes and lots
- Lot 1186591 Expires 12/31/06 Lot 1186692 Expires 12/31/06